Computrac MAX 5000XL

₹2,509,824.00

Availability:
Lead Times Vary by Product
Adding to cart… The item has been added

The Computrac® MAX® 5000XL moisture, solids, and ash analyzer can provide accurate readings from a single sample, offering many of the same features of thermogravimetric analyzers at a fraction of the cost. With a maximum temperature of 600°C and a maximum sample size of 100 grams, the MAX® 5000XL has the versatility for a wide range of material-testing tasks. 
 

  • Versatile: Get accurate readings for both ash and moisture content from a single sample.
  • Fast: Moisture and ash readings in half an hour or less.
  • Simple: One-button operation with no intermediate weighing or cooling steps required.
  • Durable: Made in the USA with an all-steel enclosure, the MAX 5000XL was designed for heavy-duty use.
  • Accurate: Correlates to popular methods, including ASTM D5630, ASTM D4574, ASTM E1534 and more.

Features and Benefits

  • High-Temperature Testing: The heater of the Computrac® MAX® 5000XL can reach temperatures as high as 600ºC, giving it the ability to perform both loss on drying (LOD) and loss on ignition (LOI) testing and making it an ideal alternative to expensive, thermogravimetric analyzers.
  • Rugged Construction: Equipped with a ruggedly designed metal casing and a high-temperature heater, the MAX 5000XL is designed to provide lab-quality data whether it is in the lab or on the production floor.
  • Versatile: The MAX 5000XL is able to test material up to 100 times faster than standard reference methods and is ideal for everything from plastics and pharmaceuticals to foods, biomass materials, and more.
  • Flexible Ending Criteria: User-adjustable ending criteria help to optimize test results and performance for your unique application or material.
  • Real-Time Results: The MAX 5000XL offers simple, menu-driven operation and a user-programmable interface with a keypad and large color display that allows users to view real-time moisture curve and rate of moisture loss graphs during testing.
  • Multistage Testing: Several tests can be linked in order to form a single, multistage test that can change temperatures, landing criteria, and times between each test segment. This allows the instrument to test for moisture, solids, and ash content with a single sample.
  • Self-Cleaning Oven: The oven of the MAX 5000XL features a self-cleaning cycle that simplifies routine maintenance. It runs for 45 minutes at 550ºC.
  • Web Server: The optional web server allows users to monitor tests remotely and check-in with operators. It also lets users download results and calibration reports, view the audit log, and transfer programs between instruments.

Specifications

Moisture | Solids Range:

0.1% to 99.9%

Ash | LOI Range:

0.5% to 100%

Moisture Resolution:

0.0001%

Balance Resolution:

0.0001 g

Moisture Repeatability:

<5% RSD for samples >10% moisture

Ash Repeatability:

≤5% RSD for samples >10% ash

Temperature Range:

25°C to 600°C

Heating Element:

Ceramic

Sample Size:

100 mg to 100 g

Results:

Moisture, Solids, Dry Weight, Ash, LOI

Ending Criteria:

User-adjustable: Prediction, Rate, Time, Reliability and 4 other combinations

Memory:

Stores up to 250 programs, last 1000 test
results and last 100 graphs

Statistical Analysis:

Mean, SD, RSD

Balance Calibration:

Menu-driven calibration by the end-user;
NIST traceable calibration performed by the manufacturer

Heater Calibration:

Menu-driven, NIST traceable calibration by the end-user with optional
Temperature Calibration Kit

Operating Environment:

0-35 ºC at <50% RH;
0-30 ºC at <80% RH

Power Requirements:

100-120 VAC, 50/60 Hz, 8A or
220-240 VAC, 50/60 Hz, 4A

Dimensions:

18.5” L x 12.7” W x 9.5” H
(47 cm L x 32.3 cm W x 24 cm H)

Weight:

31 lbs. (14 kg)

Warranty:

Two years, factory parts and labor (one year international)

Rear Panel Connections:

Ethernet, serial port, parallel port

Display:

1/4 VGA, 320 x 240 pixel, color

Certifications:

UL, CE
FCC Part 15, Class A Tested to standards EN55022,
EN 90101-1 Temperature Calibration Interface certified for use
with the MAX® 5000XL

ASTM & Regulatory Compliances

21 CFR Part 11 Compliant (optional): Meet regulatory compliance standards for pharmaceutical and medical device companies
ASTM D6980-12:
 Standard Test Method for Determination of Moisture in Plastics by Loss in Weight (MAX® 4000XL and MAX® 5000XL)

ASTM D7232-06: Standard Test Method for Rapid Determination of the Nonvolatile Content of Coatings by Loss in Weight (MAX® 4000XL and MAX® 5000XL)
ASTM C471M-16: Standard Test Methods for Chemical Analysis of Gypsum and Gypsum Products (Metric)

It also correlates well with popular methods, including ASTM D5630, ASTM D4574, ASTM E1534, and more.

Learn how you can try a Computrac® for free. 

Contact our sales team for more information on our Computrac® series. 

Downloads

 Computrac MAX 5000XL Brochure